Thermo Gravimetric Analyser

FullSizeRender 28 1 1547634831 65283

The STA 8000 simultaneous thermal analyzer offers performance, reliability and productivity.

This provides the analysis of sample heated by a small furnace (~20 cc volume) capable of operating from 15 to 1000 degrees Celsius. The STA sensor is a double pan differential temperature sensor which is calibrated to generate
data for heat flow to the sample with an accuracy of 5% or better. The weight sensor is located remotely below the
furnace where it is isolated from sample decomposition products by inert purge through a narrow channel. This provides microgram-level weight change detectability of sample loss or oxidative gain. Unless otherwise noted the purge gas used for this note was nitrogen at a flow rate of 100 cc/min. The use of argon instead of nitrogen is supported in the Pyris™ software which controls the flow rate of gas through the furnace chamber and provides for gas switching and flow rate changes.
  • Simultaneous analysis of TG with DTA mode for fast enhanced result interpretation
  • Wide temperature range allow measurements from below room temperature to 1000 °C. Superior accuracy and sensitivity in a small package with an attractive price
  • Superior weight, heat flow and temperature accuracy in a small package with an attractive price
  • Top loading balance for ease of sample loading
  • Balance below furnace for optimum isolation from contamination
  • Vertical displacement balance sensor (not rotational) provides weight insensitivity to sample position, e.g., no spurious weight change from melting

 

Contact Us

CENTRAL SOPHISTICATED INSTRUMENTATION FACILITY (CSIF)
University of Calicut
Thenhipalam PO
Malappuram - 673635 PIN
 
Copyright © 2018 University of Calicut. All Rights Reserved | Powered By CUCC