Bruker's MultiRAM is a stand-alone high-performance Fourier transform Raman spectrometer. When sample fluorescence is a problem, FT-Raman analysis with near-infrared excitation at 1064nm is the only solution. As sample fluorescence can be orders of magnitude more intense than Raman scattering, the presence of fluorescence often precludes observation of Raman scattering. The MultiRAM has a large sample compartment to utilise an extensive range of pre-aligned sampling accessories featuring Bruker ́s automatic accessory recognition. The new Raman video stage makes Raman analysis a breeze. The MultiRAM can accommodate a second laser and detection system for use with 785nm.
SPECIFICATIONS
| Laser Type | λ = 1064 nm P max ≤ 0.5, 1.0 W Divergence angle: 2 mrad |
|
Resolution |
Standard: better than 0.8 cm -1 |
| Spectral range | 3600 - 70 cm -1 Stokes shift |
| Wavenumber accuracy | 0.1 cm -1 |
| Collection optics | High-throughput 180° collection lens standard |
| Detector | InGaAs (room temperature, high sensitivity) |
| Interferometer | Mechanical ROCKSOLID interferometer, permanently aligned |
| Sampling | Pre-aligned sample stage with PC-controlled z-positioning |
| Source | • White light source for Raman background correction • Built-in alignment lamp for sample alignment and calibration • Laser |
| software | OPUS |